                    SLD Offline Processing Report for
                      Prepared by:  JLH from SLDWWW
                           23 Jul 1996 07:34:56

                                     --------Monitor Evaluation---------

     Run # Polar      ZSLD    #LUM   -RTH--    -Ev DSP-    -CDC HV- -Overall-
------------------------------------------------------------------------
    35357  .7426      161     362     (1)        ok          ok       ok
    35358    0        184     375     (2)        ok          ok       bad
    35359    0        12      26      (3)        ok          ok       bad
    35375  .7581      54      98      (4)        ok         91%       ?
    35376  .7442      34      78      (5)        *          96%       ?
------------------------------------------------------------------------

(1) run 35357:  bad pol 27%
                DTFD not flat

(2) run 35358:  polarization missing on AIS with flag message on AIS
                DTFD not flat

(3) run 35359:  polarization missing on AIS but no  flag message appears
                on AIS page
                no muon filter passes; zero correlation matrix diag element
                DTFD not flat

(4) run 35375:  bad pol = 36 %
                cdc on = 91%
                muon filter and diag element correlation matrix = 0
                definite slope to DCHD plot
                DTFD also has structure
                several ev disp events with layers off in cdc

(5) run 35376:  several ev disp events with layers off in cdc
                sharp break in DELD plot


Looking in the Zchkday I see structure in the VEP VECT and Vertex
detector efficiency plots VECT and VEPH. I think this is maybe serious.